VibrioTox | On-line Broadband Toxicity Analyzer

VibrioTox®

On-line Broadband Toxicity analyzer

 

The VibrioTox®  is a true on-line and continuous 24/7 broadband toxicity monitor. This genuine on-line system deploys the light emitting bacteria Vibrio Fischeri. Planktonic Vibrio Fischeri are found in very low quantities in almost all oceans of the world, preferentially found in temperate and sub-tropical waters. By using these living organisms AppliTek was able to develop a unique instrument that can reliably monitor the biological impact of contamination on water bodies. The Vibrio Fischeri bacterium emits light when in a sound condition. VibrioTox® samples continuously the water body and brings the water sample for a fixed amount of time in contact with the bacteria culture. A decrease in light output means that the bacteria are being intoxicated, which is processed by the incorporated software.

Early Warning System:  the VibrioTox®  on-line analyzer is set up to detect more than 2,500 toxins but can basically respond to any possible contaminant both organic or inorganic, metals and metal compounds.


Watch a trailer movie on this analyzer system (0:48 min.):

  Vibriotox trailer


Applications

  • Security of drinking water: intake water, (interconnected) distribution systems
  • River water monitoring
  • Industrial discharges (compliance)
  • Control of water recycling

 

General features

Analysis method compliant with standard methods for detection and measurement of toxicity in water

Human interface: 5.7" color touch screen with function buttons. Toxicity is expressed as % INHIBITION, either as a graph or as values. Can remotely be taken over by LAN ethernet software (VNC).

Standard auto-validation, auto-cleaning (desinfectant)

Moving parts: smooth action cleaning, drain and sample pumps with silicone tubing

Incorporated industrial PC with result & alarm data export (1000 results including sample stream, date, time)

Multifunctional sealed USB port on left side of the housing (result download, program upload, reporting)

Communication: active 4 - 20 mA, Ethernet, RS232, TCP-IP

 

Compliance

The luminescent bacteria method used in the VibrioTox® has been validated and authorized by NEN, EN and ISO, as a Water Quality-Determination of the Inhibitory Effect of Water Samples on the Light Emission of Vibrio Fischeri.

 

Options / Extra

Analysis consumable packs for 1 month continuous operation.

Autosampler taking additional samples for laboratory analysis in case of positive alarms.

This system can be part of a total monitoring solution for river or drinking water applications; please check the detailed section on this.

» Go to  Industries & Utilities /  Water utility

 

 

 VibrioTox® On-line Broadband Toxicity analyzer

 

Images